First experimental results of time-of-flight reconstruction on an LSO PET scanner
Creators
- 1. CPS Innovations, 810 Innovation Drive, Knoxville, TN 37919 (United States)
Description
Time-of-flight (TOF) positron emission tomography (PET) was studied and preliminarily developed in the 1980s, but the lack of a scintillator able to deliver at the same time proper time resolution and stopping power has prevented this technique from becoming widespread and commercially available. With the introduction of LSO in PET, TOF is now a feasible option. TOF reconstruction has been implemented in the CPS Hi-Rez PET scanner, both with 2D filtered-back-projection (FBP2D) and 3D ordered subset expectation maximization (OSEM3D). A new procedure has been introduced in the time alignment to compensate for the limited digital time resolution of the present electronics. A preliminary version of scatter correction for TOF has been devised and is presented. The measured time resolution of 1.2 ns (FWHM) allowed for a signal-to-noise ratio increase of about 50% in phantoms of about 40 cm transaxial size, or a gain larger than 2 in noise equivalent counts (NEC). TOF reconstruction has shown the expected improvement in SNR, both in simulation and experimental data. First experimental results show two improvements of TOF reconstruction over conventional (non-TOF) reconstruction: a lower noise level and a better capability to resolve structures deep inside large objects
Availability note (English)
Available online at http://stacks.iop.org/0031-9155/50/4507/pmb5_19_006.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/Additional details
Identifiers
- URL
- http://stacks.iop.org/0031-9155/50/4507/pmb5_19_006.pdf; http://www.iop.org/;
- DOI
- 10.1088/0031-9155/50/19/006;
- PII
- S0031-9155(05)92987-4;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 50
- Journal Issue
- 19
- Journal Page Range
- p. 4507-4526
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 36098676
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALIGNMENT; CORRECTIONS; NOISE; PHANTOMS; POSITRON COMPUTED TOMOGRAPHY; SIGNAL-TO-NOISE RATIO; SIMULATION; STOPPING POWER; TIME RESOLUTION; TIME-OF-FLIGHT METHOD
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; EMISSION COMPUTED TOMOGRAPHY; MOCKUP; RESOLUTION; STRUCTURAL MODELS; TIMING PROPERTIES; TOMOGRAPHY