Published 1984 | Version v1
Report

Data coding in a three-dimensional neutron diffraction experiment

Description

The system of data re-coding of red-out from position-sensitive detectors in a three-dimensional neutron diffraction experiment is briefly described. For the realization of the system the method of coordinate conversion of each event from the ''laboratory'' system into coordinates of crystal back space is applied. Coordinate conversion occurs during data acquisition process. The system consists of three units in the CAMAC standard: time coder unit, wire numbers coding unit and re-coding unit. This system permits data re-coding without losses in data acquisition speed which extends the possibilities of the use of two-coordinate detectors for investigating neutron diffraction on monocrystals

Additional details

Additional titles

Original title (Russian)
Кодирование информации в трехмерном дифракционном эксперименте

Publishing Information

Imprint Title
11. International symposium on nuclear electronics
Journal Page Range
p. 139-141.
Report number
JINR-D--13-84-53

Conference

Title
11. International symposium on nuclear electronics.
Dates
6-12 Sep 1983.
Place
Bratislava (Czechoslovakia).

INIS

Country of Publication
USSR
Country of Input or Organization
USSR
INIS RN
16046563
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CAMAC SYSTEM; DATA ACQUISITION SYSTEMS; DATA PROCESSING; DIGITAL CIRCUITS; FLOWSHEETS; POSITION SENSITIVE DETECTORS
Descriptors DEC
DIAGRAMS; ELECTRONIC CIRCUITS; INFORMATION; MEASURING INSTRUMENTS; RADIATION DETECTORS

Optional Information

Notes
Imprint:11. Mezhdunarodnyj simpozium po yadernoj ehlektronike.