Published July 1980 | Version v1
Journal article

The influence of preparation parameters on the reliability of the thyroxine RIA kit RK-12

Creators

  • 1. Magyar Tudomanyos Akademia Izotop Intezete, Budapest

Description

Neither significant differences (up to 50%) in the volume of the sample, of the antiserum or of the diluting buffer, nor changes in the duration or temperature of incubation, in the age of the radioactive tracer (125I-T4), and in the mode of the elimination of the supernatant after centrifugation (suction or decantation) influence the results, provided the alterations concern both the standard and the test sera. However, all these factors cause serious errors if the standard and test sera are prepared under different conditions. (L.E.)

Additional details

Additional titles

Original title (Hungarian)
Az RK-12 tiroxin RIA keszlet erzekenysege a keszlet feldolgozasi parametereinek valtozasara

Publishing Information

Journal Title
Izotoptechnika
Journal Volume
23
Journal Issue
4
Series
Izotoptechnika.
Journal Page Range
214-228
ISSN
0004-7201

Optional Information

Notes
5 refs.