Effects of internal and external scatter on the build-up characteristics of Monte Carlo calculated absorbed dose for electron irradiation

  • 1. Hefei University of Technology, Hefei (China). Physics and Mathematics Department
  • 2. Hefei University of Technology, Hefei (China). Adult Education Department
  • 3. Anhui Province Hospital, Hefei (China). Radiation Oncology Department

Description

The effects of internal and external scatter on surface, build-up and depth dose characteristics simulated by Monte Carlo code EGSnrc for varying field size and SSD for a 10 MeV monoenergetic electron beam with and without an accelerator model are extensively studied in this paper. In particular, sub-millimetre surface PDD was investigated. The percentage depth doses affected significantly by the external scatter show a larger build-up dose. A forward shifted Dmax depth and a sharper fall-off region compared to PDDs with only internal scatter considered. The surface dose with both internal and external scatter shows a marked decrease at 110 cm SSD, and then slight further changes with the increasing SSD since few external scattered particles from accelerator model can reach the phantom for large SSDs. The sharp PDD increase for the 5 cm x 5 cm field compared to other fields seen when only internal scatter is considered is significantly less when external scatter is also present. The effect of external scatter on surface PDD is more pronounced for large fields than small fields (5 cm x 5 cm field)

Additional details

Publishing Information

Journal Title
Australasian Physical and Engineering Sciences in Medicine
Journal Volume
28
Journal Issue
3
Journal Page Range
p. 165-171
ISSN
0158-9938
CODEN
AUPMDI

Optional Information

Notes
20 refs., 9 figs.