Published September 1979 | Version v1
Journal article

Size determination of airborne plutonium by autoradiography using intensifier screen

  • 1. Power Reactor and Nuclear Fuel Development Corp., Tokai, Ibaraki (Japan). Tokai Works

Description

An autoradiographic technique in estimating size of air-borne plutonium was studied by using ZnS(Ag) intensifier screen and Polaroid film. Effects of screen thickness, a weight percent of binder added and a particle size if ZnS(Ag) power on a spot size of autoradiograph were examined. A calibration curve between alfa activity and spot size of autoradiograph produced under optimum exposure condition is presented. All procedures of the autoradiography can be performed without darkroom by the use of an improved Polaroid camera which has a double light-tight attachment. (author)

Additional details

Publishing Information

Journal Title
Hoken Butsuri
Journal Volume
14
Journal Issue
3
Series
Hoken Butsuri.
Journal Page Range
163-170
ISSN