Published March 1992
| Version v1
Journal article
Evaluation of a 3D reconstruction algorithm for multi-slice PET scanners
Creators
- 1. California Univ., Los Angeles, CA (United States). School of Medicine
Description
A fully 3D reconstruction algorithm based on filtered backprojection was evaluated for the reconstruction of data obtained with multi-slice positron emission tomography (PET) scanners which have had the septa removed. This algorithm uses forward-projection through the reconstructed images of a 2D subset of the data to complete the 3D dataset thus satisfying the condition of shift invariance. This is followed by 3D filtered backprojection. Axial sampling was doubled by combining adjacent polar angles, thus improving reconstructed axial resolution. The algorithm was tested using real and simulated datasets and gave high quality reconstructions without artifacts over a wide range of imaging conditions. (author)
Additional details
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 37
- Journal Issue
- 3
- Series
- Phys. Med. Biol.
- Journal Page Range
- 779-790
- ISSN
- 0031-9155
- CODEN
- PHMBA
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 23043044
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; EVALUATION; IMAGE PROCESSING; POSITRON COMPUTED TOMOGRAPHY; SPATIAL RESOLUTION
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; RESOLUTION; TOMOGRAPHY
Optional Information
- Contract/Grant/Project number
- Contract DE-FC03-87ER 60615; Grant NIHR01-MH37916; P01-NS15654