ISO 4037:2019 Validation of radiation qualities by means of half-value layer and Hp(10) dosimetry
- 1. Individual Monitoring Service, Helmholtz Zentrum Muenchen - German Research Center for Environmental Health -GmbH-, Otto-Hahn-Ring 6, 81739 Muenich (Germany)
Description
Radiation qualities were characterised and validated at the fully automated X-ray calibration facility of the Individual Monitoring Service at Helmholtz Zentrum Muenchen by using half-value layer and Hp(10) dosimetry approaches specified in the updated ISO 4037:2019 standard. As the ISO 4037 contains a somewhat vague description of the half-value layer procedure, we extended it to be more constrained and thus less subjective in its implementation. We specify both the measurement and data analysis steps performed in order to provide reproducible half-value layer results and compare the results with the Hp(10) dosimetry-based validation approach specified in the ISO 4037 as an alternative for field validation. Finally, we discuss the implications of our results and the extended procedure on validation and the recent changes to the ISO 4037. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncz185Additional details
Identifiers
- DOI
- 10.1093/rpd/ncz185;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 187
- Journal Issue
- 4
- Journal Page Range
- p. 438-450
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 51108204
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; DATA ANALYSIS; DOSIMETRY; IMPLEMENTATION; PERSONNEL MONITORING; RADIATION MONITORING; RADIATION QUALITY; VALIDATION
- Descriptors DEC
- DATA PROCESSING; MONITORING; PROCESSING; RADIATION MONITORING; TESTING
Optional Information
- Notes
- 6 refs.