Published February 1983
| Version v1
Report
Restricted
Application of a semi-empirical Ge(Li) detector calibration technique for large-volume samples
Description
A previously developed semi-empirical method for the calculation of the absolute peak efficiency of coaxial Ge(Li) detectors for large volume samples is reviewed and applied using a computer model and experimental determinations of calibration parameters. The efficiencies calculated for extended air, lucite and aluminum source materials are compared with those obtained with spiked sample calibrations and point source calibrations. The errors are within +-5% in most cases. For the calibrations, two versions of computer programs written in FORTRAN are presented
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS, PC A03/MF A01; 1 as DE83012801.
Files
Additional details
Publishing Information
- Imprint Pagination
- 36 p.
- Report number
- DOE/EML--409
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 14788445
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; COMPUTER CODES; EFFICIENCY; FORTRAN; GE SEMICONDUCTOR DETECTORS; LI-DRIFTED GE DETECTORS; MATHEMATICAL MODELS; S CODES; STATISTICS
- Descriptors DEC
- LI-DRIFTED DETECTORS; MATHEMATICS; MEASURING INSTRUMENTS; PROGRAMMING LANGUAGES; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS