Published November 1984 | Version v1
Journal article

Performance characteristics of X-ray tubes and generators

  • 1. Newcastle upon Tyne General Hospital (UK). Regional Medical Physics Dept.

Description

Some aspects of the performance of 132 X-ray tubes and generators used in diagnostic radiology are reported. Measurements were made for single and three phase systems of a variety of manufacturers. These include 1) effective peak tube voltage, 2) X-ray tube output and reproducibility, 3) exposure time and 4) half value layer and total filtration. For all aspects of performance examined, no significant overall difference between manufacturers was seen. (U.K.)

Additional details

Publishing Information

Journal Title
Radiography
Journal Volume
50
Journal Issue
594
Series
Radiography.
Journal Page Range
245-248
ISSN
0033-8281

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
16021269
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S07: ISOTOPES AND RADIATION SOURCES;
Descriptors DEI
PERFORMANCE TESTING; X-RAY EQUIPMENT; X-RAY TUBES
Descriptors DEC
ELECTRON TUBES; EQUIPMENT; TESTING