Published April 1994
| Version v1
Journal article
The effects of humidity on the response of radiochromic film dosimeters FWT-60-00 and Gafchromic-DM-1260
Creators
- 1. Atomic Energy of Canada Ltd., Pinawa, MB (Canada). Whiteshell Labs.
Description
The response characteristics of the radiochromic film dosimeters, nylon-based FWT-60-00 and coated polyester Gafchromic-DM-1260, were investigated as a function of relative humidity (RH) of air with which they were equilibriated, in the range of 0-94% RH. Studies were made for both γ-ray (dose rate = 2 Gy/s) and 10-MeV electron-beam (pulse dose rate = 106 Gy/s) irradiations. The response of the FWT dosimeters, unlike that of the Gafchromic, shows significant variation with RH. These results are in disagreement with the results in the recent articles. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Physics and Chemistry (1993)
- Journal Volume
- 43
- Journal Issue
- 4
- Journal Page Range
- p. 407-409.
- ISSN
- 0969-806X
- CODEN
- RPCHDM
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 25026131
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COLORIMETRIC DOSEMETERS; DOSE RATES; ELECTRON DOSIMETRY; GAMMA DOSIMETRY; HUMIDITY; MEV RANGE 01-10; RESPONSE FUNCTIONS
- Descriptors DEC
- DOSEMETERS; DOSIMETRY; ENERGY RANGE; FUNCTIONS; MEASURING INSTRUMENTS; MEV RANGE; MOISTURE