We report the first accuracy evaluation of NRC-FCs2, an atomic fountain clock developed at the National Research Council Canada. The short term stability at high atomic density is 1.1 x 10-13 at 1 second of averaging time. The typical overall fractional type B uncertainty is 2.3 x 10-16. GNSS-based frequency comparisons of NRC-FCs2 with the SI second, as reported by the BIPM over fourteen months, give a fractional frequency difference of - 2.4(3.7) x 10-16. (authors)
Availability note (English)
Available from doi: https://doi.org/10.1088/1681-7575/ab7c54Additional details