Published August 1995 | Version v1
Journal article

Pulse-height-spectrum distortion in xenon gaseous detectors for soft x rays: Experimental results

  • 1. Univ. of Coimbra (Portugal). Physics Dept.

Description

The results of an experimental determination of the pulse-height distortion in soft x-ray spectra in gaseous xenon detectors are presented. The study confirms previous Monte Carlo simulation results and emphasizes the importance of the electric field in the drift region in reducing spectral degradation while the effect of a strong magnetic field in the same region is shown to be negligible

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
42
Journal Issue
4Pt1
Journal Page Range
p. 611-614.
ISSN
0018-9499
CODEN
IETNAE

Conference

Title
medical imaging conference.
Acronym
Nuclear science symposium
Dates
30 Oct - 5 Nov 1994.
Place
Norfolk, VA (United States).

Optional Information