Accurate in vivo dosimetry of a randomized trial of prostate cancer irradiation

Description

Purpose: To guarantee an accurate dose delivery, within ±2.5%, in a Phase III randomized trial of prostate cancer irradiation (68 vs. 78 Gy) by means of a comprehensive in vivo dosimetry program. Methods and Materials: Prostate patients are generally treated in our clinic with a 3-field isocentric technique: an 8-MV anteroposterior beam and 2 18-MV wedged laterals. All fields are shaped conformally to the PTV. Patients were randomized between two dose levels of 68 Gy and 78 Gy. During treatment, the entrance and exit dose were measured for each patient with diodes. Special 2.5-mm thick steel build-up caps were applied to make the diodes appropriate for measurements in 18-MV photon beams as well. Portal images were used to verify the correct position of the diodes and to detect and correct for gas filling in the rectum that may influence the exit dose reading. Entrance and exit dose measurements were converted to midplane dose, which was used in combination with a depth dose correction to obtain the dose at the specification point. An action level of 2.5% was applied. Results: The added build-up for the diodes in the 18-MV beams resulted in correction factors that were only slightly sensitive to changes in beam setup and comparable to the corrections used in the 8-MV beams for diodes without extra build-up. The calibration factor increased almost linearly with cumulative dose: 0.7%/kGy for the 8-MV and 1.2%/kGy for the 18-MV photon beams. The introduction of average correction factors made the analysis easier, while keeping the accuracy within acceptable limits. In a period of 3 years, 225 patients were analyzed, from which 8 patients needed to be corrected. The average ratio of measured and prescribed dose was 1.009 (standard deviation [SD] 0.012) for the total group treated on two linear accelerators. When the results were analyzed per accelerator, the ratios were 1.002 (SD, 0.001) for Accelerator A and 1.015 (SD, 0.001) for Accelerator B. This difference could be attributed to the cumulative effect of three small imperfections in the performance of Accelerator B that were well within the limits of our quality assurance program. Conclusion: Diodes can be used for accurate in vivo dosimetry during prostate irradiation in high-energy photon beams. The dose delivery in this randomized trial is guaranteed within the 2.5% limits on an individual patient basis. This could not be achieved without the in vivo dosimetry program, despite our high-standard quality assurance program of treatment delivery

Additional details

Identifiers

PII
S036030160001556X;

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
49
Journal Issue
5
Journal Page Range
p. 1409-1418
ISSN
0360-3016
CODEN
IOBPD3

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
35010446
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
ACCURACY; CARCINOMAS; DOSIMETRY; IN VIVO; LOCAL IRRADIATION; PROSTATE; QUALITY ASSURANCE; RADIOTHERAPY
Descriptors DEC
BODY; DISEASES; GLANDS; IRRADIATION; MALE GENITALS; MEDICINE; NEOPLASMS; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY

Optional Information

Copyright
Copyright (c) 2001 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.