SU-F-T-578: Characterization of Vidar DosimetryPro Advantage RED Scanner with Application to SBRT and SRS QA
Description
Purpose: To use Gafchromic EBT3 film to quantify key dosimetric characteristics of the Vidar DosimetryPro Advantage RED film scanner for use in SBRT/SRS QA, by analyzing scanner uniformity and dose sensitivity. Method: Gafchromic EBT3 film was used in this study. Films were irradiated using 6MV FFF and 10MV FFF beams from a Varian Edge linear accelerator, with setup of 100cm SAD at depth 5 cm. Nine doses were delivered per film, with calibration dose ranges of 1–10 Gy and 3–24 Gy for 6MV FFF, and 3–27 Gy for 10MV FFF. Films were scanned with the long side of the film parallel to the detector array. Dose calibration curves were fitted to a 3rd degree polynomial. The derivative of a calibration curve was taken to determine the scanner's sensitivity per dose delivered (OD/Gy). Scanner non-uniformity was calculated in 2 dimensions by taking the mean of standard deviation in each row and column. Absolute dose SRS/SBRT Gamma analyses were performed with passing criteria of 3% and 1mm DTA. For comparison, Gamma analyses were also performed using an Epson Expression 10000 XL. Results: Uniformity for the Vidar scanner was 0.37% +/− 0.03% in the perpendicular to scan direction and 0.67% +/− 0.05% in the parallel to scan direction, with an overall uniformity of 0.52% +/− 0.03%. Epson red channel uniformity was 0.85% +/− 0.05% and 0.88% +/− 0.08% for the green channel. The Vidar average dose sensitivity from 1–10 Gy was 0.112 OD/Gy and 0.061 OD/Gy for 3–24 Gy. SBRT/SRS Gamma pass rates were 97.8 +/− 1.4 for Vidar and 97.5 +/− 1.4 for Epson. Conclusion: The Vidar scanner has 41% less non-uniformity compared to Epson XL10000 green channel. The dose sensitivity is 2–3 time greater for the Vidar scanner compared to the Epson in the SRS/SBRT dose range of 5–24 Gy.
Additional details
Identifiers
- DOI
- 10.1118/1.4956763;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 43
- Journal Issue
- 6
- Journal Page Range
- vp.
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48097811
- Subject category
- S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- CALIBRATION; DIFFERENTIAL THERMAL ANALYSIS; LINEAR ACCELERATORS; RADIATION DOSES; SENSITIVITY
- Descriptors DEC
- ACCELERATORS; DOSES; THERMAL ANALYSIS
Optional Information
- Notes
- (c) 2016 American Association of Physicists in Medicine