Published December 2004 | Version v1
Report Open

Development of a gas filled two-dimensional position sensitive x-ray detector for small angle scattering

Description

A two-dimensional position-sensitive X-ray detector was developed for small angle X-ray scattering experiments. It has a 190x190 mm2 effective window and can get 0.3 mm spatial resolution using Ar mixture gases. The fabricated detector was tested at 4C1 SAXS beamline at Pohang Accelerator Laboratory for small angle X-ray scattering. It would play an important role in study of polymer and protein structure. It could be possible to realize rapid time resolved experiments(<0.1s) with fast data acquisition system. Because the detector is noise free in comparison to CCD detector, it has a great advantage in solution scattering which has weak scattering power

Availability note (English)

Available from INIS in electronic form; Also available from Korea Atomic Energy Research Institute, Taejon (Korea, Republic of)

Files

36095461.pdf

Files (10.1 MB)

Name Size Download all
md5:9f22f1a825a0182f3a55c860ad09794b
10.1 MB Preview Download

Additional details

Publishing Information

Imprint Pagination
83 p.
Report number
KAERI/TR--2870/2004

INIS

Optional Information

Notes
18 refs, 38 figs, 2 tabs