Published June 1970
| Version v1
Report
Time response, linearity, and gain of 250-micron silicon solid-state detectors
Description
Experiments are described in which the time response, linearity, and gain were determined for the 250-micron silicon solid-state detectors used in a solid-state magnetic gamma-ray spectrometer. Time-response measurements performed at the EG and G/AEC linear accelerator (linac) determined detector impulse response, step response, frequency-domain amplitude, and phase as a function of frequency. Linearity measurements performed with the linac determined detector current and charge linearity at bias levels up to 500 V. Gain measurements made in the counting mode at the NBS Dynamitron (0.25 to 1.0 MeV) and Van de Graaff (2.0 to 4.0 MeV) accelerators determined gain as a function of incident electron energy. (U.S.)
Availability note (English)
MF available from INIS under the Report Number.Additional details
Publishing Information
- Imprint Pagination
- 105 p.
- Report number
- EGG--1183-2239
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 7224061
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COMPTON SPECTROMETERS; GAIN; GAMMA SPECTROMETERS; PERFORMANCE TESTING; RESPONSE FUNCTIONS; SI SEMICONDUCTOR DETECTORS; SURFACE BARRIER DETECTORS; TIMING PROPERTIES
- Descriptors DEC
- FUNCTIONS; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SPECTROMETERS; TESTING