Published May 2018 | Version v1
Miscellaneous

Calculation of Most Efficient Volume of Detection by Using the Differential Volume Method in the EXVol Code

  • 1. Seoul National University, Seoul (Korea, Republic of)
  • 2. Korea Atomic Energy Research Institute, Daejeon (Korea, Republic of)

Description

To calculate the full energy (FE) absorption peak efficiency for arbitrary volume sample, we developed and verified the EXVol (Efficiency calculator for eXtended Voluminous source) code which is based on effective solid angle method. Correction techniques for factors that interfere with γ-ray spectrum analysis are required for reliable nuclear characterization and radioactivity determination. And in the measurement of voluminous source of various forms and media, it is necessary to calibrate the effects of the self-attenuation and coincidence summing of the γ-ray depending on the volume and density of the sample. In this study, the performance of the EXVol code was extended to obtain the detection efficiency distribution at a specific position in the volume sample. And we obtained γ-ray detection efficiency and determined the most efficient volume contributing to detection efficiency of the large voluminous Si ingot source. As a result of the study, γ-ray detection efficiency and radiation distribution were obtained in the large volume and high density medium source.

Part of:
Proceedings of the KNS 2018 Spring Meeting

Additional details

Identifiers

Publishing Information

Publisher
KNS
Imprint Place
Daejeon (Korea, Republic of)
Imprint Title
Proceedings of the KNS 2018 Spring Meeting
Imprint Pagination
vp.
Journal Page Range
[2 p.]

Conference

Title
2018 Spring Meeting of the KNS
Dates
16-18 May 2018
Place
Jeju (Korea, Republic of)

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
50059660
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
COMPUTER CALCULATIONS; CORRECTIONS; DATA; DETECTION; E CODES; EFFICIENCY; HIGH-PURITY GE DETECTORS; PERFORMANCE; VERIFICATION
Descriptors DEC
COMPUTER CODES; GE SEMICONDUCTOR DETECTORS; INFORMATION; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS

Optional Information

Notes
2 refs, 2 figs, 1 tab