Influence on dose calculation by difference of dose calculation algorithms in stereotactic lung irradiation. Comparison of pencil beam convolution (inhomogeneity correction: batho power law) and analytical anisotropic algorithm
Creators
- 1. Kyushu Univ., Hospital, Fukuoka, Fukuoka (Japan)
Description
The monitor unit (MU) was calculated by pencil beam convolution (inhomogeneity correction algorithm: batho power law) [PBC (BPL)] which is the dose calculation algorithm based on measurement in the past in the stereotactic lung irradiation study. The recalculation was done by analytical anisotropic algorithm (AAA), which is the dose calculation algorithm based on theory data. The MU calculated by PBC (BPL) and AAA was compared for each field. In the result of the comparison of 1031 fields in 136 cases, the MU calculated by PBC (BPL) was about 2% smaller than that calculated by AAA. This depends on whether one does the calculation concerning the extension of the second electrons. In particular, the difference in the MU is influenced by the X-ray energy. With the same X-ray energy, when the irradiation field size is small, the lung pass length is long, the lung pass length percentage is large, and the CT value of the lung is low, and the difference of MU is increased. (author)
Additional details
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi
- Journal Volume
- 65
- Journal Issue
- 8
- Journal Page Range
- p. 1064-1072
- ISSN
- 0369-4305
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 41009227
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; ANISOTROPY; CALCULATION METHODS; CARCINOMAS; CAT SCANNING; COMPARATIVE EVALUATIONS; FRACTIONATED IRRADIATION; LINEAR ACCELERATORS; LUNGS; METASTASES; PATIENTS; PLANNING; RADIATION DOSES; RADIATION MONITORING; RADIOTHERAPY
- Descriptors DEC
- ACCELERATORS; BODY; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DISEASES; DOSES; EVALUATION; IRRADIATION; MATHEMATICAL LOGIC; MEDICINE; MONITORING; NEOPLASMS; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; RESPIRATORY SYSTEM; THERAPY; TOMOGRAPHY