A miniature X-ray tube is tested in place of a conventional radioisotope source. The test is made for the resistive plastic tube counter whose cathode induced signals are utilized for position measurement. The optimum resolution of 0.17 mm is obtained by collimated X-rays. This result is promising for designing a precision luminosity monitor of the AMY detector at TRISTAN, KEK. (orig.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods Phys. Res., Sect. A
- Journal Volume
- 255
- Journal Issue
- 3
- Series
- Nucl. Instrum. Methods Phys. Res., Sect. A.
- Journal Page Range
- 477-481
- ISSN
- 0168-9002
- CODEN
- NIMAE
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 18057878
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BEAM LUMINOSITY; BEAM MONITORS; CATHODES; ORGANIC POLYMERS; POSITION SENSITIVE DETECTORS; SPATIAL RESOLUTION; TESTING; TRISTAN STORAGE RINGS; TUBES; X-RAY DETECTION; X-RAY SOURCES; X-RAY SPECTRA; X-RAY TUBES
- Descriptors DEC
- DETECTION; ELECTRODES; ELECTRON TUBES; EQUIPMENT; MEASURING INSTRUMENTS; MONITORS; ORGANIC COMPOUNDS; POLYMERS; RADIATION DETECTION; RADIATION DETECTORS; RADIATION SOURCES; RESOLUTION; SPECTRA; STORAGE RINGS; X-RAY EQUIPMENT