Published May 2000 | Version v1
Journal article

Test-method for determination of Mo(6) with phenylfluorone on a solid phase of fibrous sorbents

  • 1. RAN, Inst. Geokhimii i analiticheskoj khimii im. V.I. Vernadskogo, Moscow (Russian Federation)

Description

A system of phenylflorone - fibrous sorbent filled with an AV-17 anion exchanger is studied for visual determination of Mo. Mo(6) is sorbed from a solution of ascorbic acid on a solid phase and then detected with phenylfluorone in a 0.5M HCl. Within the limits of a chromatic scale 0.025 - 0.4 μg/ml change in color (from orange to magenta) is observed. The test method is approved on river and tap water. The limits of Mo determination is 0.02 mg/liter and the relative error of determination does not exceed 25 %. A procedure duration is 10 min

Abstract (Russian)

Изучена система фенилфлуорон - волокнистый сорбент, наполненный анионообменником АВ-17, для визуального определения молибдена. Mo(6) сорбируют из раствора аскорбиновой кислоты на твердой фазе и детектируют раствором фенилфлуорона в 0.5 М HCl. В пределах цветовой шкалы 0.025 - 0.4 мкг/мл наблюдается изменение окраски из оранжевой в малиновую. Тест-метод апробирован на речной и водопроводной воде. Предел обнаружения молибдена составляет 0.02 мг/л. Относительная погрешность не превышает 25 %. Продолжительность анализа - 10 минут

Additional details

Additional titles

Original title (Russian)
Тест-метод определения молибдена (6) с фенилфлуороном на твердой фазе волокнистых сорбентов

Publishing Information

Journal Title
Zavodskaya Laboratoriya
Journal Volume
66
Journal Issue
5
Journal Page Range
p. 13-15
ISSN
0321-4265
CODEN
ZVDLAU

Optional Information

Notes
10 refs., 2 figs., 1 tab.