Modelling the imaging performance and low contrast detectability in digital mammography
- 1. Department of Medical Physics, Medical School, University of Patras, 265 00 Patras, Greece, (Greece)
- 2. Department of Medical Instruments Technology, Technological Educational Institution of Athens, Aigaleo, 122 10 Athens (Greece)
Description
A digital x-ray mammography is a modern method for the early detection of breast cancer. The quality of a mammography image depends on various factors, the detector structure and performance being of primary importance. The aim of this work was to develop an analytical model simulating the imaging performance of a new commercially available digital mammography detector. This was achieved within the framework of the linear cascaded systems (LCS) theory. System analysis has allowed the estimation of important image quality metrics such as the Modulation Transfer Function (MTF), the Noise Power Spectrum (NPS) and the Detective Quantum Efficiency (DQE). The detector was an indirect detection system consisting of a large area, 100μm thick, CsI:TI scintillator coupled to an active matrix array of amorphous silicon (a-Si:H) photodiodes combined with thin film transistors (TFT). Pixel size was 100μm, while the active pixel dimension was 70μm. MTF and DQE data were calculated for air kerma conditions of 25, 53, 67 μGy using a 28 kVp Mo-Mo x-ray spectrum. The theoretical results were compared with published experimental data. The deviation between the theoretical and experimental MTF curves was less than 4%, while the DQE differences were found at an acceptable level. The model was also used to estimate system's capability to detect low contrast objects in the breast. It was estimated that, in the breast gland, low contrast structures larger than 1.4mm can be adequately identified by the above system.
Availability note (English)
Available from http://dx.doi.org/10.1088/1748-0221/4/06/P06004Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Instrumentation
- Journal Volume
- 4
- Journal Issue
- 06
- Journal Page Range
- p. P06004
- ISSN
- 1748-0221
Conference
- Title
- 4. international conference on imaging technologies in biomedical sciences: from medical images to clinical information - Bridging the gap
- Acronym
- ITBS07
- Dates
- 22-28 Sep 2007
- Place
- Milos Island (Greece)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41012096
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- AIR; BIOMEDICAL RADIOGRAPHY; DIAGRAMS; MAMMARY GLANDS; PERFORMANCE; PHOTODIODES; QUANTUM EFFICIENCY; SILICON; THIN FILMS; X RADIATION; X-RAY SPECTRA
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; EFFICIENCY; ELECTROMAGNETIC RADIATION; ELEMENTS; FILMS; FLUIDS; GASES; GLANDS; INFORMATION; IONIZING RADIATIONS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIATIONS; RADIOLOGY; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; SEMIMETALS; SPECTRA