Published 1996 | Version v1
Journal article

The nice high-energy neutron facility: dosimetry intercomparisons

  • 1. Centre de Lutte Contre le Cancer Antoine Lacassagne, 06 - Nice (France)
  • 2. Universite Catholique de Louvain (UCL), Louvain-la-Neuve (Belgium)

Description

Neutron dosimetry intercomparison studies have been undertaken at the Nice neutron-therapy facility with the staff at Louvain-la-Neuve which has had wide experience in both dosimetric and radiobiological intercomparisons. Tissue equivalent (TE) ionization chambers were first calibrated in 60Co beams and then exposed in the neutron beam at different depths in a water phantom; the largest difference observed in neutron beam measurements with all the chambers tested was 0.89 %, and most of them agreed to within less than 0.5 %. The gamma component at four depths was derived from measurements with Geiger-Mueller counters; the results obtains with the two counters (Nice and Louvain-la-Neuve), expressed as a percentage of the dose (neutron + gamma), agreed to within less that 0.03 % and the value increased from 1.4 to 4 % between 2 and 20 cm in depth. (author)

Additional details

Additional titles

Original title (French)
Intercomparaison dosimetrique sous le faisceau de neutrons a Nice

Publishing Information

Journal Title
Bulletin du Cancer. Radiotherapie
Journal Volume
83
Series
3 refs.
Journal Page Range
p. 153-156.
ISSN
0924-4212
CODEN
BCRAEE

Conference

Title
Meeting of the European heavy particle therapy group.
Dates
4-6 May 1995.
Place