Published December 1976 | Version v1
Journal article

Image aberrations in an imperfectly-shaped electrostatic analyser

  • 1. Stichting voor Fundamenteel Onderzoek der Materie, Amsterdam (Netherlands). Instituut voor Atoom en Molecuulfysica

Description

Image aberrations caused by mechanical imperfections of the electrodes of an electrostatic energy analyser are calculated. Deviations of the electrodes from the ideal cylindrical shape may be of any type. In examples, the following cases are treated: mis-positioning (eccentricity and non-parallelism) and mis-shaping (conical, hyperbolic and elliptic) of the electrodes, as well as the general case. First, the electric field is calculated in the imperfect analyser, and then the deviations of the ion orbits in this distorted field from the corresponding orbits in the ideal electrostatic analyser. This new method permits the setting of exact values for the tolerances in the manufacture and adjustment of the electrodes. (Auth.)

Additional details

Identifiers

Publishing Information

Journal Title
International Journal of Mass Spectrometry and Ion Physics
Journal Volume
22
Journal Issue
3-4
Series
Int. J. Mass Spectrom. Ion Phys.
Journal Page Range
259-271
ISSN
0020-7381

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent