Determination of nominal accelerating potential
Creators
- 1. Department of Radiation Oncology, University of Massachusetts Medical Center, Worcester, Massachusetts 01605
Description
We present a simple linear relationship between the nominal accelerating potential (NAP) and the ratios of ionization measurements made with constant source--detector distance and at two different phantom thicknesses. This relationship can be used as a standard, unambiguous method for determining NAP for use in dosimetry and quality control