Published November 1986 | Version v1
Journal article

Determination of nominal accelerating potential

  • 1. Department of Radiation Oncology, University of Massachusetts Medical Center, Worcester, Massachusetts 01605

Description

We present a simple linear relationship between the nominal accelerating potential (NAP) and the ratios of ionization measurements made with constant source--detector distance and at two different phantom thicknesses. This relationship can be used as a standard, unambiguous method for determining NAP for use in dosimetry and quality control

Additional details

Publishing Information

Journal Title
Med. Phys.
Journal Volume
13
Journal Issue
6
Series
Med. Phys.
Journal Page Range
961-962
ISSN
0094-2405
CODEN
MPHYA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
18036533
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
CALIBRATION STANDARDS; DOSIMETRY; IONIZING RADIATIONS; PHANTOMS; QUALITY CONTROL; THICKNESS
Descriptors DEC
CONTROL; DIMENSIONS; MOCKUP; RADIATIONS; STRUCTURAL MODELS