Purpose: To present an empirically-based beam model for computing dose deposited by kilovoltage (kV) x-rays and validate it for radiographic, CT, CBCT, superficial, and orthovoltage kV sources. Method and Materials: We modeled a wide variety of imaging (radiographic, CT, CBCT) and therapeutic (superficial, orthovoltage) kV x-ray sources. The model characterizes spatial variations of the fluence and spectrum independently. The spectrum is derived by matching measured values of the half value layer (HVL) and nominal peak potential (kVp) to computationally-derived spectra while the fluence is derived from in-air relative dose measurements. This model relies only on empirical values and requires no knowledge of proprietary source specifications or other theoretical aspects of the kV x-ray source. To validate the model, we compared measured doses to values computed using our previously validated in-house kV dose computation software, kVDoseCalc. The dose was measured in homogeneous and anthropomorphic phantoms using ionization chambers and LiF thermoluminescent detectors (TLDs), respectively. Results: The maximum difference between measured and computed dose measurements was within 2.6%, 3.6%, 2.0%, 4.8%, and 4.0% for the modeled radiographic, CT, CBCT, superficial, and the orthovoltage sources, respectively. In the anthropomorphic phantom, the computed CBCT dose generally agreed with TLD measurements, with an average difference and standard deviation ranging from 2.4 ± 6.0% to 5.7 ± 10.3% depending on the imaging technique. Most (42/62) measured TLD doses were within 10% of computed values. Conclusions: The proposed model can be used to accurately characterize a wide variety of kV x-ray sources using only empirical values

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
41
Journal Issue
8
Journal Page Range
p. 3
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes
(c) 2014 American Association of Physicists in Medicine