Comparison of entrance exposure and signal-to-noise ratio between an SBDX prototype and a wide-beam cardiac angiographic system
Creators
- 1. Department of Medicine and Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin 53792 (United States)
- 2. Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin 53706 (United States)
- 3. NexRay, Inc., Los Gatos, California 95030 (United States)
Description
The scanning-beam digital x-ray (SBDX) system uses an inverse geometry, narrow x-ray beam, and a 2-mm thick CdTe detector to improve the dose efficiency of the coronary angiographic procedure. Entrance exposure and large-area iodine signal-to-noise ratio (SNR) were measured with the SBDX prototype and compared to that of a clinical cardiac interventional system with image intensifier (II) and charge coupled device (CCD) camera (Philips H5000, MRC-200 x-ray tube, 72 kWp max). Phantoms were 18.6-35.0 cm acrylic with an iohexol-equivalent disk placed at midthickness (35 mg/cm2 iodine radiographic density). Imaging was performed at 15 frame/s, with the disk at mechanical isocenter and an 11-cm object-plane field width. The II/CCD system was operated in cine mode with automatic exposure control. With the SBDX prototype at maximum x-ray output (120 kVp, 24.3 kWp), the SBDX SNR was 107%-69% of the II/CCD SNR, depending on phantom thickness, and the SBDX entrance exposure rate was 10.7-9.3 R/min (9.4-8.2 cGy/min air kerma). For phantoms where an equal-kVp imaging comparison was possible (≥23.3 cm), the SBDX SNR ranged from 47% to 69% of the II/CCD SNR while delivering 6% to 9% of the II/CCD entrance exposure rate. From these measurements it was determined that the relative SBDX entrance exposure at equal SNR would be 31%-16%. Results were consistent with a model for relative entrance exposure at equal SNR, which predicted a 3-7 times reduction in entrance exposure due to SBDX's comparatively low scatter fraction (5.5%-8.1% measured, including off-focus radiation), high detector detective quantum efficiency (66%-73%, measured from 70 to 120 kVp), and large entrance field area (1.7x-2.3x, for the same object-plane field width). With improvements to the system geometry, detector, and x-ray source, SBDX technology is projected to achieve conventional cine-quality SNR over a full range of patient thicknesses, with 5-10 times lower skin dose
Additional details
Identifiers
- DOI
- 10.1118/1.2198198;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 33
- Journal Issue
- 8
- Journal Page Range
- p. 2728-2743
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38026362
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; CDTE SEMICONDUCTOR DETECTORS; CHARGE-COUPLED DEVICES; CORONARIES; DOSIMETRY; IMAGE INTENSIFIERS; IMAGES; IOHEXOL; KERMA; PATIENTS; PHANTOMS; QUANTUM EFFICIENCY; RADIATION DOSES; SIGNAL-TO-NOISE RATIO; SKIN; THICKNESS; X RADIATION; X-RAY SOURCES; X-RAY TUBES
- Descriptors DEC
- ARTERIES; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; CONTRAST MEDIA; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; DIMENSIONS; DOSES; EFFICIENCY; ELECTROMAGNETIC RADIATION; ELECTRON TUBES; EQUIPMENT; IONIZING RADIATIONS; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIATION DETECTORS; RADIATION SOURCES; RADIATIONS; RADIOLOGY; SEMICONDUCTOR DETECTORS; SEMICONDUCTOR DEVICES; STRUCTURAL MODELS; X-RAY EQUIPMENT
Optional Information
- Notes
- (c) 2006 American Association of Physicists in Medicine