Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies

Description

An investigation into the effect of pressure, temperature and humidity in air on photon fluence at a typical mammography, low bremsstrahlung energy (25 kVp), has been carried out. Pressure values corrected for humidity at varying temperatures were employed to determine the density of moist air. Using the corrected moist air densities, the X-ray photon fluence remaining after air attenuation has been computed assuming an X-ray focal spot-detector distance of 65 cm. Bremsstrahlung spectral distributions and the effects of pressure, temperature and humidity on the photon fluence from molybdenum and tungsten targets are illustrated. Comparative results suggest that such effects could be significant and need to be considered when calculating exposure levels from low-energy photons. The investigation showed that air kerma values from an X-ray spectrum that has significant lower-energy components is likely to be more sensitive to changes in pressure, temperature and humidity than the air kerma from an X-ray spectrum with lower-energy components less pronounced. Non-negligible air kerma values are involved

Additional details

Additional titles

Augmented title (English)
X-ray photon fluence; Attenuation; Mammography; Bremsstrahlung; Air kerma

Identifiers

DOI
10.1016/S0969-806X(03)00398-0;
arXiv
arXiv:hep-th/9802018v1;
PII
S0969806X03003980;

Publishing Information

Journal Title
Radiation Physics and Chemistry (1993)
Journal Volume
68
Journal Issue
5
Journal Page Range
p. 707-720
ISSN
0969-806X
CODEN
RPCHDM

Optional Information

Copyright
Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.