Investigation of applicability of alanine and radiochromic detectors to dosimetry of proton clinical beams
- 1. National Inst. of Standards and Technology (NML), Gaithersburg, MD (United States). Ionizing Radiation Div.
- 2. Gosudarstvennyj Komitet po Ispol'zovaniyu Atomnoj Ehnergii SSSR, Moscow (Russian Federation). Inst. Teoreticheskoj i Ehksperimental'noj Fiziki
Description
Cancer therapy studies using proton accelerators are underway in several major medical centers in the U.S., Russia, Japan and elsewhere. To facilitate dosimetry intercomparisons between these laboratories, alanine-based detectors produced at the National Institute of Standards and Technology and commercially available radiochromic films were studied for their possible use as passive transfer dosimeters for clinical proton beams. Evaluation of characteristics of these instruments, including the LET dependence of their response of proton energy, was carried out at the Institute of Theoretical and Experimental Physics. Results of absolute dose measurements were regarded as a preliminary step of dose intercomparison between ITEP and NIST. Measurements made in a number of experiments showed average agreement between the ITEP and NIST dosimetry standards to 2.5%. (Author)
Additional details
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 46
- Journal Issue
- 12
- Journal Page Range
- p. 1355-1362.
- ISSN
- 0969-8043
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 27022226
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- CALIBRATION STANDARDS; CHEMICAL DOSEMETERS; EVALUATION; INTERLABORATORY COMPARISONS; PHOTOGRAPHIC FILM DOSEMETERS; PROTON DOSIMETRY; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY
- Descriptors DEC
- DOSEMETERS; DOSIMETRY; MEASURING INSTRUMENTS; MEDICINE; STANDARDS; THERAPY