Published December 1995 | Version v1
Journal article

Investigation of applicability of alanine and radiochromic detectors to dosimetry of proton clinical beams

  • 1. National Inst. of Standards and Technology (NML), Gaithersburg, MD (United States). Ionizing Radiation Div.
  • 2. Gosudarstvennyj Komitet po Ispol'zovaniyu Atomnoj Ehnergii SSSR, Moscow (Russian Federation). Inst. Teoreticheskoj i Ehksperimental'noj Fiziki

Description

Cancer therapy studies using proton accelerators are underway in several major medical centers in the U.S., Russia, Japan and elsewhere. To facilitate dosimetry intercomparisons between these laboratories, alanine-based detectors produced at the National Institute of Standards and Technology and commercially available radiochromic films were studied for their possible use as passive transfer dosimeters for clinical proton beams. Evaluation of characteristics of these instruments, including the LET dependence of their response of proton energy, was carried out at the Institute of Theoretical and Experimental Physics. Results of absolute dose measurements were regarded as a preliminary step of dose intercomparison between ITEP and NIST. Measurements made in a number of experiments showed average agreement between the ITEP and NIST dosimetry standards to 2.5%. (Author)

Additional details

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
46
Journal Issue
12
Journal Page Range
p. 1355-1362.
ISSN
0969-8043
CODEN
ARISEF