Published April 15, 2004 | Version v1
Report

Omega Dante Soft X-Ray Power Diagnostic Component Calibration at the National Synchrotron Light Source

Description

The Dante soft x-ray spectrometer installed on the Omega laser facility at the Laboratory for Laser Energetics, University of Rochester is a twelve-channel filter-edge defined x-ray power diagnostic. It is used to measure the absolute flux from direct drive, indirect drive (hohlraums) and other plasma sources. Calibration efforts using two beam lines, U3C (50eV-1keV) and X8A (1keV-6keV) at the National Synchrotron Light Source (NSLS) have been implemented to insure the accuracy of these measurements. We have calibrated vacuum x-ray diodes, mirrors and filters

Availability note (English)

Available from http://www.llnl.gov/tid/lof/documents/pdf/306962.pdf; PURL: https://www.osti.gov/servlets/purl/15014117-8mADl3/native/

Additional details

Publishing Information

Imprint Pagination
vp.
Report number
UCRL-CONF--203618

Conference

Title
15. Topical Conference on High Temperature Plasma Diagnostics
Dates
19-22 Apr 2004
Place
San Diego, CA (United States)

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
39025933
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S43: PARTICLE ACCELERATORS;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ACCURACY; CALIBRATION; LASERS; MIRRORS; NSLS; X-RAY SPECTROMETERS
Descriptors DEC
MEASURING INSTRUMENTS; RADIATION SOURCES; SPECTROMETERS; SYNCHROTRON RADIATION SOURCES

Optional Information

Contract/Grant/Project number
W-7405-ENG-48
Notes
PDF-FILE: 8 ; SIZE: 0.3 MBYTES
Funding organization
US Department of Energy (United States)