Published 1989 | Version v1
Journal article

Indirect spectrophotometric determination of BIDA, DISIDA, DTPA and MDP in labelled compounds

  • 1. Rand Afrikaans Univ., Johannesburg (South Africa). Dept. of Chemistry
  • 2. Atomic Energy Corp. of South Africa (Pty) Ltd., Pretoria (South Africa)

Description

N-(4-(n-butyl)acetanilide)iminodiacetic acid (BIDA), N-(2,6-diisopropylacetanilide)iminodiacetic acid (DISDA), diethylenetriaminepentaacetic acid (DTPA) and methylene diphosphonic acid (MDP) are used in labelling kits. The contents of BIDA, DISIDA or MDP of the 99mTc-labelled compounds can be determined (indirectly) spectrophotometrically with copper, eriochrome cyanine R (ECC) and dodecyl-ethyldimethylammonium bromide (DEDA) in a sodium barbital buffered system at pH 8.5. The calibration curves obey Beer's Law from 0 to 40 μg/25 mL for BIDA and DISIDA, 0 to 60 μg/25 mL for DTPA and 0 to 100 μg/10 mL for MDP. (author)

Additional details

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
40
Journal Issue
6
Series
Appl. Radiat. Isot.
Journal Page Range
525-529
CODEN
ARISE