Published May 2011 | Version v1
Journal article

Silicon photomultipliers for positron emission tomography detectors with depth of interaction encoding capability

  • 1. Thunder Bay Regional Health Science Centre, Thunder Bay, ON (Canada)
  • 2. Department of Medical Biophysics, University of Toronto, Toronto, ON (Canada)
  • 3. Department of Physics, Lake Head University, Thunder Bay, ON (Canada)

Description

Silicon photomultipliers (SiPMs) are receiving increasing attention in the field of positron emission tomography (PET) detectors. Compared to photomultiplier tubes, they offer novel detector configurations for the extraction of depth of interaction (DOI) information, or enable emerging medical imaging modalities such as simultaneous PET-magnetic resonant imaging (MRI). In this article, we used 2x2x20 mm3 LYSO scintillator crystals coupled to SiPMs on both ends (dual-ended readout configuration) to evaluate the detector performance for DOI-PET applications. We investigated the effect of scintillator crystal surface finishing on sensitivity and resolution of DOI, as well as on energy and timing resolution. Measurements indicate DOI sensitivity and resolution of 7.1% mm-1 and 2.1±0.6 mm for saw-cut, and 1.3% mm-1 and 9.0±1.5 mm, for polished scintillator crystals, respectively. Energy resolution varies from 19% when DOI is in the center, to 15% with DOI at either end of the saw-cut crystal, while it remains constant at ∼14% for polished scintillators. Based on our results we conclude that 2x2x20 mm3 saw-cut (without any special side wall polishing) LYSO crystals coupled to 2x2 mm2 silicon photomultipliers are optimal for isotropic 2 mm resolution DOI-PET applications.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2010.06.180

Additional details

Identifiers

DOI
10.1016/j.nima.2010.06.180;
PII
S0168-9002(10)01370-7;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
633
Journal Issue
Suppl.1
Journal Page Range
p. S250-S254
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
11. international workshop on radiation imaging detectors
Acronym
IWORID
Dates
28 Jun - 2 Jul 2009
Place
Prague (Czech Republic)

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
43053985
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CRYSTALS; ENERGY RESOLUTION; GAMMA RADIATION; PERFORMANCE; PHOTOMULTIPLIERS; POSITRON COMPUTED TOMOGRAPHY; READOUT SYSTEMS; SENSITIVITY
Descriptors DEC
COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; EMISSION COMPUTED TOMOGRAPHY; IONIZING RADIATIONS; PHOTOTUBES; RADIATIONS; RESOLUTION; TOMOGRAPHY