Performance of a microcalorimeter with a superconducting transition edge thermometer
Description
The performance of a single-pixel microcalorimeter is described and analyzed. The microcalorimeter has a Cu absorber which is isolated from the heat bath by means of a SixNy membrane and is connected to a voltage-biased Ti/Au proximity bilayer thermometer with a transition temperature of about 150 mK. The measured FWHM energy resolution for 5.9 keV X-ray photons is 12.4 eV for this sensor while 2.36(kBT2C)1/2 equals 9.1 eV. The detector time constant is about 1 ms. The measured current-voltage characteristic and current noise spectra are analyzed in detail in the context of the energy resolution. High frequency excess noise is measured in the superconducting transition and is briefly discussed
Additional details
Identifiers
- PII
- S016890029901356X;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 444
- Journal Issue
- 1-2
- Journal Page Range
- p. 192-195
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Portugal
- INIS RN
- 35021135
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CRYOGENICS; ENERGY RESOLUTION; NOISE; RADIATION DETECTION; SUPERCONDUCTING DEVICES; SUPERCONDUCTIVITY; SUPERCONDUCTORS; TEMPERATURE RANGE 0000-0013 K; X-RAY DETECTION; X-RAY SPECTROSCOPY
- Descriptors DEC
- DETECTION; ELECTRIC CONDUCTIVITY; ELECTRICAL PROPERTIES; PHYSICAL PROPERTIES; RADIATION DETECTION; RESOLUTION; SPECTROSCOPY; TEMPERATURE RANGE
Optional Information
- Copyright
- Copyright (c) 2000 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.