Published February 2011 | Version v1
Journal article

Long-term neutron dose fading study of LiF: Mg,Cu,P TLD

  • 1. Uniformed Services University of the Health Sciences, Bethesda, MD 20814 (United States)
  • 2. Naval Dosimetry Center, 8901 Wisconsin Avenue, Bethesda, MD 20889 (United States)

Description

An extensive study of neutron dose fading in LiF:Mg,Cu,P thermoluminescent dosimeters (TLD) using commercially-available dosimeters and equipment for their processing was undertaken. During the 52 week study, three thousand TLDs were stored for various lengths of time before and after being exposed to a plutonium beryllium radiation source. The TLDs were subsequently processed and the resulting doses were compared to the reference exposure. Both a loss of signal and a loss of sensitivity were evaluated. The results of this study have shown that the commercially produced LiF:Mg,Cu,P TLD has no statistically significant change in sensitivity or change in signal with up to 52 weeks of pre-irradiation or post-irradiation time. The results of this study will provide the technical basis for increasing the exposure record accuracy, and extending the usable lifetime of dosimeters.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.radmeas.2010.11.012

Additional details

Identifiers

DOI
10.1016/j.radmeas.2010.11.012;
PII
S1350-4487(10)00374-4;

Publishing Information

Journal Title
Radiation Measurements
Journal Volume
46
Journal Issue
2
Journal Page Range
p. 199-204
ISSN
1350-4487
CODEN
RMEAEP

INIS

Optional Information

Copyright
Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.