Estimation of patient dose in abdominal CT examination in some Sudanese hospitals

  • 1. Sudan Academy of Sciences (SAS) Atomic Energy Of Council, Kartoum (Sudan)

Description

The use of CT in medical diagnosis delivers radiation doses to patients that are higher than those from other radiological procedures. The aim of this study was to estimate radiation doses in abdomen CT examinations of patients in two Sudanese hospitals. Details were obtained from approximately 80 CT examinations and included all age groups ( adults and pediatric). The results from the two hospitals were compared with each other as well as with the IAEA guidance level for this particular investigation. The estimation of radiation doses were carried out by calculating volume dose index (CTD1vol), dose length product (DLP), doses to some organs of interest and effective dose (E) using the software program CT EXPO V2.1. The study showed that the mean DLP of the one hospitals ASH is 1736.7 mGy.cm which is by far much higher than that for the other hospital NMDC which stands at 185.3 mGy.cm, as well as higher than the IAEA level which is 696 mGy.cm. The study showed that the mean CTD1vol for patients in ASH is 36.2 mGy which again higher than that for the other hospital which is 3.9 mGy and higher than the IAEA level which is 10.9 mGy calculating the effective dose for patients in the two hospitals reveals that the mean effective dose of patient in one hospital (ASH) is 26.25 mSv, which is quite high compared with other hospital (NMDC), which has the mean value of 2.8 mGv and also higher than the IAEA level from this investigation which is 7.6 mSv. Regarding organ doses, the study showed that organ doses in hospital ASH are always higher than that calculated in hospital NMDC and the highest doses in both hospital were delivered to the kidneys with mean values of 50.24 mGy and 5045 mGy for the two hospitals respectively. The study showed that there is an urgent need for optimizing patient doses in such CT examinations. This can be ensured by providing training and retraining for workers and conducting quality control measurements and preventive maintenance regularly so as to detect any flaws early enough before they negatively affect the image, which may necessitate re-imaging and then increase patients dose. (Author)

Availability note (English)

Also availoble at Sudan Atomic Energy Commission

Files

48039603.pdf

Files (3.6 MB)

Name Size Download all
md5:9c7d1a88c6702b1e72ed5fd1243aab77
3.6 MB Preview Download

Additional details

Publishing Information

Imprint Place
Khartoum (Sudan)
Imprint Pagination
54 p.
Report number
INIS-SD--639

Optional Information

Notes
33 refs., 12 figs., 4 tabs.