Published November 2013 | Version v1
Journal article

Study of size, type and the kerma approximation for thermoluminescent dosimetry

  • 1. Ferdowsi University of Mashhad, Mashhad (Iran, Islamic Republic of). Physics Department, School of Sciences

Description

This study aimed to investigate effects of the thermoluminescent dosimeter (TLD) size and type on the absorbed dose value by using of Monte Carlo calculations. The options in creating conditions to establish the kerma approximation were also studied. The Monte Carlo N-Particle (MCNPX 2.4.0) transport code was used to design simulations. Results of this work indicate that if common mineral materials of TLDs are replaced by air and a huge volume is applied for the TLD, the accurate assessment of absorbed doses is possible while the photon energy fluence in the TLD cell is convoluted with mass energy absorption coefficients of the real TLD material. In this method the simulation run-time is strongly decreased. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
298
Journal Issue
2
Journal Page Range
p. 813-819
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
23 refs.