Published October 1986 | Version v1
Report Open

A transuranic aerosol measurement system: Preliminary results

Description

We have completed the design, fabrication, and assembly of a computer-based prototype system for the measurement of transuranic aerosols in the workplace and environment. This system (called WOTAMS for Workplace Transuranic Aerosol Measurement System) incorporates two detectors: (1) an in-line solid-state alpha detector that sends out an alarm the moment a transuranic release occurs, and (2) an in-vacuum detector that increases off-line-analysis sensitivity. The in-line sensitivity of the system is better than 5.0 MPC-h, and the in-vacuum sensitivity exceeds 0.5 MPC-h. 5 refs., 8 figs., 1 tab

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A02/MF A01; 1 as DE87002389.

Files

18056005.pdf

Files (162.3 kB)

Name Size Download all
md5:ad8c3c1706ee7c6ed2b7b264ef40d880
162.3 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
7 p.
Report number
UCRL--95481

Conference

Title
Nuclear science and nuclear power systems symposium.
Dates
29-31 Oct 1986.
Place
Washington, DC (USA).

Optional Information

Notes
Portions of this document are illegible in microfiche products.
Secondary number(s)
CONF-861007--19.