The photon calibration of a tissue-equivalent ionisation chamber for neutron dosimetry
Description
The poor agreement in the photon calibrations of tissue-equivalent ionisation chambers between different participants in two major neutron dosimetry intercomparisons is noted. The methods of converting measurements made with a calibrated secondary standard exposure meter to the absorbed dose in the wall of a tissue-equivalent chamber are discussed. Calibrations of a TE ionisation chamber were made in five photon beams (250 kV, 300 kV, 137Cs, 60Co and 4 MV) and the different calibration factors were compared both for in-air and in-phantom measurements. Good agreement was found in these factors provided that allowance was made for the attenuation and scattering in the wall and build-up cap of the TE chamber for the in-air measurements, and that revised values of Csub(lambda) were used for the in-phantom measurements at the two higher energies. (author)
Additional details
Publishing Information
- Journal Title
- Phys. Med. Biol.
- Journal Volume
- 25
- Journal Issue
- 2
- Series
- Phys. Med. Biol.
- Journal Page Range
- 215-224
- ISSN
- 0031-9155
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 11562750
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- CALIBRATION; COMPARATIVE EVALUATIONS; ENERGY DEPENDENCE; ERRORS; GAMMA RADIATION; IONIZATION CHAMBERS; NEUTRON DOSIMETRY; PHOTONS; RADIOTHERAPY; RELIABILITY; TISSUE-EQUIVALENT DETECTORS; X RADIATION
- Descriptors DEC
- DOSIMETRY; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; IONIZING RADIATIONS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; MEDICINE; RADIATION DETECTORS; RADIATIONS; THERAPY