Critical angle laser refractometer

  • 1. Instituto de Investigaciones Electricas, Avenida Reforma 113, Colonia Palmira, Temixco, 62490 Morelos (Mexico)
  • 2. Centro de Ciencias Fisicas, UNAM, Avenida Universidad s/n, Cuernavaca, 62251 Morelos (Mexico)
  • 3. Blackett Laboratory, Imperial College, Prince Consort Road, SW7 2BZ London (United Kingdom)

Description

A simple laser refractometer based on the detection of the critical angle for liquids is presented. The calibrated refractometer presents up to 0.000 11 of uncertainty when the refractive index is in the range between 1.300 00 and 1.340 00. The experimental setup is easy to construct and the material needed is available at most optics laboratories. The calibration method is simple and can be used in other devices. The refractive index measurements in aqueous solutions of sodium chloride were carried out to test the device and a linear dependence between the refractive index and the salt concentration was found

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
77
Journal Issue
3
Journal Page Range
p. 035101-035101.4
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2006 American Institute of Physics