Published November 30, 2011 | Version v1
Journal article

Influence of fabrication procedure on the electrochemical performance of Ag/AgCl reference electrodes

  • 1. Department of Biomedical and Inorganic Chemistry, Laboratoire National de Metrologie et d'Essais, 1 Rue Gaston Boissier, 75015 Paris (France)
  • 2. Analytical Science Division, National Physical Laboratory, Teddington, Middlesex TW11 0LW (United Kingdom)

Description

The influence of several parameters in the preparation procedure of thermal-electrolytic Ag/AgCl electrodes on the resulting electrode performance has been studied. In particular, we report the effect on electrode performance of subtle variations in the preparation of silver oxide paste used for electrode manufacture, in thermal annealing conditions employed and in the procedure for electrochemically converting a fraction of the electrode from silver to silver chloride. Scanning electron microscopy and electrochemical impedance spectroscopy have been used to study the characteristics of the electrodes produced. This work reveals a correlation between the electrochemical behaviour and surface physical characteristics - in particular electrode porosity. The outputs of this study have positive implications for improving the accuracy and comparability of primary pH measurement.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.electacta.2011.08.089

Additional details

Identifiers

DOI
10.1016/j.electacta.2011.08.089;
PII
S0013-4686(11)01336-3;

Publishing Information

Journal Title
Electrochimica Acta
Journal Volume
56
Journal Issue
27
Journal Page Range
p. 10009-10015
ISSN
0013-4686
CODEN
ELCAAV

Optional Information

Copyright
Copyright (c) 2011 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.