Published June 1, 1988 | Version v1
Journal article

A simple phoswich system

  • 1. Southampton Univ. (UK). Dept. of Physics

Description

Normal phoswich detector systems use a combination of NaI(Tl) and CsI(Na) scintillators and require the application of careful pulse-shape discriminator techniques to resolve the two components in the scintillation light output which have decay constants of 250 and 630 ns respectively. These techniques provide a good anticoincidence veto efficiency for a relatively narrow range in the ratio of energy deposits in the two crystals and for a detector system whose temperature is carefully controlled. This paper describes the performance of a simple phoswich which makes use of the fast UV signal from a BaF2 crystal to provide a prompt veto signal. The performance to be expected from various combinations of a BaF2 anticoincidence crystal with other primary detectors is presented. These simulations have been verified by simple experimental tests. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res., Sect. A
Journal Volume
269
Journal Issue
1
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
230-236
ISSN
0168-9002
CODEN
NIMAE