Dosimetric variances anticipated from breathing- induced tumor motion during tomotherapy treatment delivery
Creators
- 1. Department of Radiation Oncology, Washington University School of Medicine, St. Louis, MO 63110 (United States)
Description
In their classic paper, Yu et al (1998 Phys. Med. Biol. 43 91) investigated the interplay between tumor motion caused by breathing and dynamically collimated, intensity-modulated radiation delivery. The paper's analytic model assumed an idealized, sinusoidal pattern of motion. In this work, we investigate the effect of tumor motion based on patients' breathing patterns for typical tomotherapy treatments with field widths of 1.0 and 2.5 cm. The measured breathing patterns of 52 lung- and upper-abdominal-cancer patients were used to model a one-dimensional motion. A convolution of the measured beam-dose profiles with the motion model was used to compute the dose-distribution errors, and the positive and negative dose errors were recorded for each simulation. The dose errors increased with increasing motion magnitude, until the motion was similar in magnitude to the field width. For the 1.0 cm and 2.5 cm field widths, the maximum dose-error magnitude exceeded 10% in some simulations, even with breathing-motion magnitudes as small as 5 mm and 10 mm, respectively. Dose errors also increased slightly with increasing couch speed. We propose that the errors were due to subtle drifts in the amplitude and frequency of breathing motion, as well as changes in baseline (exhalation) position, causing both over- and under-dosing of the target. The results of this study highlight potential breathing-motion-induced dose delivery errors in tomotherapy. However, for conventionally fractionated treatments, the dose delivery errors may not be co-located and may average out over many fractions, although this may not be true for hypofractionated treatments.
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/54/8/019Additional details
Identifiers
- DOI
- 10.1088/0031-9155/54/8/019;
- PII
- S0031-9155(09)93639-9;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 54
- Journal Issue
- 8
- Journal Page Range
- p. 2541-2555
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41012282
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CT-GUIDED RADIOTHERAPY; ERRORS; EXHALATION; LUNGS; MOTION; NEOPLASMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RESPIRATION; SIMULATION
- Descriptors DEC
- BODY; CLEARANCE; DISEASES; DOSES; EXCRETION; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; RADIOTHERAPY; RESPIRATORY SYSTEM; THERAPY