Published 2005 | Version v1
Journal article

A strategy to qualify the performance of radiographic monitors

  • 1. Univ. Inst. for Applied Radiophysics, Grand-Pre 1, CH-1007 Lausanne (Switzerland)

Description

The purpose of this work was to compare standard desktop display systems with dedicated medical display systems. The set of image tests proposed by the American Association of Physicists in Medicine (AAPM TG18) was used to assess a Philips 107S desktop display system and a Siemens medical display. Three observers performed the subjective assessment, in a non-concerted manner. The objective assessment was performed using a CCD camera according to the AAPM TG18 procedure. The results clearly demonstrate the inadequacy of standard desktop display systems in the framework of diagnostic radiology. Moreover, a good correlation between the subjective and objective assessment methods was obtained. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nch565

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
114
Journal Issue
1-3
Journal Page Range
p. 192-197
ISSN
0144-8420

Conference

Title
2. Malmoe Conference on Medical X-Ray Imaging - Optimisation strategies in medical X-ray imaging
Dates
23-25 Apr 2004
Place
Malmoe (Sweden)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
France
INIS RN
44035792
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CHARGE-COUPLED DEVICES; COMPUTER CODES; IMAGES; INFORMATION SYSTEMS; INTERFACES; MONITORS; PERFORMANCE; QUALITY CONTROL; RADIATION MONITORING; RADIOLOGY; SENSITIVITY; SPECIFICITY; STANDARDS
Descriptors DEC
CONTROL; MEASURING INSTRUMENTS; MEDICINE; MONITORING; NUCLEAR MEDICINE; SEMICONDUCTOR DEVICES

Optional Information

Notes
6 refs