Published 2005
| Version v1
Journal article
A strategy to qualify the performance of radiographic monitors
- 1. Univ. Inst. for Applied Radiophysics, Grand-Pre 1, CH-1007 Lausanne (Switzerland)
Description
The purpose of this work was to compare standard desktop display systems with dedicated medical display systems. The set of image tests proposed by the American Association of Physicists in Medicine (AAPM TG18) was used to assess a Philips 107S desktop display system and a Siemens medical display. Three observers performed the subjective assessment, in a non-concerted manner. The objective assessment was performed using a CCD camera according to the AAPM TG18 procedure. The results clearly demonstrate the inadequacy of standard desktop display systems in the framework of diagnostic radiology. Moreover, a good correlation between the subjective and objective assessment methods was obtained. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/nch565Additional details
Identifiers
- DOI
- 10.1093/rpd/nch565;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 114
- Journal Issue
- 1-3
- Journal Page Range
- p. 192-197
- ISSN
- 0144-8420
Conference
- Title
- 2. Malmoe Conference on Medical X-Ray Imaging - Optimisation strategies in medical X-ray imaging
- Dates
- 23-25 Apr 2004
- Place
- Malmoe (Sweden)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 44035792
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CHARGE-COUPLED DEVICES; COMPUTER CODES; IMAGES; INFORMATION SYSTEMS; INTERFACES; MONITORS; PERFORMANCE; QUALITY CONTROL; RADIATION MONITORING; RADIOLOGY; SENSITIVITY; SPECIFICITY; STANDARDS
- Descriptors DEC
- CONTROL; MEASURING INSTRUMENTS; MEDICINE; MONITORING; NUCLEAR MEDICINE; SEMICONDUCTOR DEVICES
Optional Information
- Notes
- 6 refs