Published 1992 | Version v1
Journal article

Determination of effective backscatter volume

Creators

  • 1. Troxler Electronics Labs., Research Triangle Park, NC (United States)

Description

An attempt is made to estimate the effective volume of a density gauge in the backscatter mode. The first and second scatter collision densities are determined using Monte Carlo methods. Statistical estimation is used to determine the response from each collision site; from this, the response densities are determined. The response densities are used to determine the contribution each volume element makes to the effective volume of the system. The requirements for the method used to determine the effective volume are (a) that the method be reasonable in simple or obvious geometrics, (b) that the effective volume approach a limit as the choice of volume clement size decreases, and (c) that the effective volume approach a limit as the real volume becomes infinite

Additional details

Publishing Information

Journal Title
Transactions of the American Nuclear Society
Journal Volume
65
Journal Issue
1
Journal Page Range
p. 34-35.
ISSN
0003-018X
CODEN
TANSAO

Conference

Title
2. topical meeting on industrial radiation and radioisotope measurement applications.
Dates
8-11 Sep 1992.
Place
Raleigh, NC (United States).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
24072790
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BACKSCATTERING; COMPUTERIZED SIMULATION; DENSIMETERS; MEASURING METHODS; MONTE CARLO METHOD; RADIOMETRIC GAGES
Descriptors DEC
CALCULATION METHODS; MEASURING INSTRUMENTS; SCATTERING; SIMULATION

Optional Information

Secondary number(s)
CONF-920919--.