Published June 21, 2005 | Version v1
Journal article

Response of semi-insulating 100μm thick GaAs detector for α-particles, γ-rays and X-rays

  • 1. Heavy Ion Laboratory, Warsaw University, ul. Pasteura 5a, 02-093 Warsaw (Poland)
  • 2. Institute of Electronic Materials Technology, Warsaw (Poland)
  • 3. Center for Inter-Faculty Individual Studies in Mathematical and Natural Sciences, University of Warsaw, Warsaw (Poland)

Description

The 100μm thick, transmission, fully depleted GaAs Schottky barrier detector fabricated from LEC bulk GaAs crystals has been tested. The basic structure of the detector consists of a Cr/Au Schottky contact and an Au/Ge/Ni/Au alloyed ohmic contact. Pulse height spectra for α-particles resulting from irradiation with triple source 239Pu, 241Am, 244Cm, with energies 5.155, 5.486, 5.805MeV were acquired. The energy resolution for these energies was better than 10keV (1.8%). Using 241Am 5.486MeV α-particles a charge collection efficiency (CCE) of about 94% was obtained at 570V bias potential. The spectra of low-energy γ-rays 59.5keV and X-rays with average energy 16.2keV from 241Am source measured at room temperature are presented. An energy resolution of about 6.1keV (10.3%) FWHM at 59.5keV was obtained

Additional details

Identifiers

DOI
10.1016/j.nima.2005.02.015;
PII
S0168-9002(05)00757-6;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
545
Journal Issue
3
Journal Page Range
p. 716-720
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2005 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.