Survey of diagnostic X-Ray units at Recife- Brazil
Creators
- 1. Departamento de Energia Nuclear, UFPE, Recife, PE(Brazil)
Description
This paper reports reports the result of a survey aiming to determine the working conditions of x-ray units operating in hospital and clinics in the city of Recife, Pernamburgo, Brazil. The study comprised 27 x-ray units from both public and private diagnostic services operating. The following parameters were evaluated: a) coincidence between the luminous and radiation fields; b) alignment of the radiation beam; c) agreement between the real and preset values of the kVp and exposure time. The result showed that the discrepancy between the luminous and radiation fields stayed in the recommended 1 to 2% interval in 78.2% of the equipment surveyed, all units passing the alignment test (deviation of less than 3%). The results also showed that there is only a fair accuracy regarding both preset kilovoltage and exposure time, with discrepancies of more than 10% between ''true'' irradiation times(in85% of the case), and between the present and applied kilovoltage (in 46% of the inspected units)
Additional details
Additional titles
- Original title (English)
- Memorias de Seguridad Radiologica y Nuclear. 4th Congreso Regional. Congreso Regional IRPA. Vol 2
Publishing Information
- Publisher
- Centro de Informacion de La Energia Nuclear
- Imprint Place
- La Habana (Cuba)
- Imprint Title
- Proceedings of 4. regional congress on radiological and nuclear safety. Regional congress IRPA. Vol. 2
- Imprint Pagination
- 330 p.
- Journal Page Range
- p. 61-64
Conference
- Title
- 4. Regional congress on radiological and nuclear safety; Regional Congress IRPA
- Original Conference Title
- 4. Congreso Regional sobre Seguridad Radiologica y Nuclear. Congreso Regional IRPA
- Dates
- 19-23 Oct 1998
- Place
- La Habana (Cuba)
INIS
- Country of Publication
- Cuba
- Country of Input or Organization
- Cuba
- INIS RN
- 30056431
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference, Numerical Data, Non-conventional Literature
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; BRAZIL; EXPERIMENTAL DATA; HOSPITALS; ICRP; PHOTOMETERS; QUALITY CONTROL; RADIATION DOSES; RADIATION PROTECTION; X-RAY EQUIPMENT
- Descriptors DEC
- BUILDINGS; CONTROL; DATA; DEVELOPING COUNTRIES; DIAGNOSTIC TECHNIQUES; EQUIPMENT; INFORMATION; INTERNATIONAL ORGANIZATIONS; LATIN AMERICA; MEASURING INSTRUMENTS; MEDICAL ESTABLISHMENTS; MEDICINE; NUMERICAL DATA; SOUTH AMERICA
Optional Information
- Notes
- 3 figs., 3 refs. Imprint:Memorias de Seguridad Radiologica y Nuclear. 4th Congreso Regional. Congreso Regional IRPA. Vol 2