Published 1972
| Version v1
Journal article
Characteristics of proportional counters used in X-ray radiometric analysis
Description
Counters with extended (up to 10 cm2) apertres have been described and the results obtained are presented. CPM-15 and CPM-16 counters are intended for sensoring and spectral analysis of X-rays in the range of 2.5 to 23 and 2.5 to 15 keV, respectively. The only difference is in the gas composition. The CPM-15 counter is filled with 90 per cent xenon plus 10 per cent methane used as a quenching dopant. The CPM-16 is filled with 90 per cent argon and 10 per cent methane. The life time is 1010 pulses at 1800 V for CPM-16 and at 2200 V for CPM-15
Additional details
Additional titles
- Original title (Russian)
- Характеристики пропорциональных счетчиков для рентгенорадиометрического анализа
Publishing Information
- Journal Title
- Radiats. Tekh.
- Journal Issue
- no.8
- Series
- Radiats. Tekh.
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 8281485
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- AMPLITUDES; ARGON; METHANE; MIXTURES; PERFORMANCE; PROPORTIONAL COUNTERS; PULSES; RESOLUTION; SPECIFICATIONS; X-RAY EMISSION ANALYSIS; X-RAY SPECTRA; XENON
- Descriptors DEC
- ALKANES; CHEMICAL ANALYSIS; DISPERSIONS; ELEMENTS; HYDROCARBONS; MEASURING INSTRUMENTS; NONDESTRUCTIVE ANALYSIS; NONMETALS; ORGANIC COMPOUNDS; RADIATION DETECTORS; RARE GASES; SPECTRA
Optional Information
- Notes
- 6 figs.; 1 table.