Published November 1992 | Version v1
Journal article

Fundamental study of helical scanning CT

Description

We evaluated spatial resolution in the longitudinal axis with helical scanning CT using a fourth-generation fast CT scanner. We made a phantom by stringing acrylic balls (6.5 mmφ x 8 and 9 mmφ x 6). The aquired images were processed by MPR and assessed visually to evaluate axis resolution. With the conventional scanning method, the partial volume effect varied with the starting position, but helical scanning was able to reconstruct high-resolution images using continuous raw data. During helical scanning, axis resolution varied depending on the slice width and sliding speed of the couch top. Even if the sliding speed was kept constant at 4 mm/sec, axis resolution was superior with a slice width of 2 mm to that with one of 5 mm. (author)

Additional details

Additional titles

Subtitle (English)
Evaluation of spatial resolution in the longitudinal axis

Publishing Information

Journal Title
Nippon Igaku Hoshasen Gakkai Zasshi
Journal Volume
52
Journal Issue
11
Journal Page Range
p. 1540-1544.
ISSN
0048-0428
CODEN
NHGZAR

INIS