Published February 21, 1996 | Version v1
Journal article

Calibration of analyzing magnet for beam energy measurement using the recoil mass separator HIRA

  • 1. Nuclear Sci. Centre, New Delhi (India)

Description

Beam energy measurements have been carried out using the energy achromatic recoil mass separator HIRA at NSC. The energy was obtained from a measurement of the time-of-flight through HIRA which offers a flight distance of about 8.8 m. The method is quite useful for the calibration of the analyzer magnet over the entire operating range. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
370
Journal Issue
2-3
Journal Page Range
p. 315-318.
ISSN
0168-9002
CODEN
NIMAER