Published 1987
| Version v1
Journal article
Position sensitive detector for β spectrometers
- 1. AN Kazakhskoj SSR, Alma-Ata. Inst. Yadernoj Fiziki
Description
Positron sensitive detector (PSD) placed into beta spectrometer focal plane is described. The detector overall dimensions are 90x30x100 mm. To detect electrons microchannel plate (MCP) chevron is used. The distance between the plates is 0.15 mm and the applied potential difference is 250 V. A charge escaping from M CP gets to a wire delay line placed at o.5 mm distance from chevron and appearing to be an organic glass plate with 2x20 mm cross section and a gold-plated wire 28 μm in diameter, coiled to the plate at 0.1 mm pitch. PSD application increases the data setting rate in a beta spectrometer by two orders and improves 5-10 times signal/background relation
Additional details
Additional titles
- Original title (Russian)
- Позиционно-чувствительный детектор для β-спектрометров
Publishing Information
- Journal Title
- Izv. Akad. Nauk Kaz. SSR, Ser. Fiz.-Mat.
- Journal Issue
- no.2
- Series
- Izv. Akad. Nauk Kaz. SSR, Ser. Fiz.-Mat.
- ISSN
- 0002-3191
- CODEN
- IAKFB
INIS
- Country of Publication
- Kazakhstan
- Country of Input or Organization
- USSR
- INIS RN
- 19096002
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BETA SPECTROMETERS; ELECTRON DETECTION; POSITION SENSITIVE DETECTORS; SPECIFICATIONS
- Descriptors DEC
- CHARGED PARTICLE DETECTION; DETECTION; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; SPECTROMETERS