Performance analysis on SNR of scientific grade optical CCD
Creators
- 1. Research Center of Laser Fusion, CAEP, Mianyang (China)
- 2. School of Physics and Electron, Central South University, Changsha (China)
- 3. School of Information Engineering, Nanchang Hangkong University, Nanchang (China)
Description
Based on the noise source, considering that the fluctuations of incident photons, dark current and biased signal satisfy the Poisson distribution, the corrected signal-to-noise-ratio (SNR) relation is obtained. On condition that the gain, readout rate and cooling temperature of CCD are set experimentally to be 3, 100 kHz and −25 ℃ respectively, the photon transfer curve is obtained. Meanwhile, by setting different experimental conditions, the effect of the dark current, readout rate and the system gain on the SNR of the scientific grade optical CCD is discussed respectively. The experimental results are in good agreement with the theory, which provides the good experimental basis for design of the scientific grade optical streak camera and building of the automated laboratory platform used for weak light calibration. (authors)
Additional details
Publishing Information
- Journal Title
- Nuclear Fusion and Plasma Physics
- Journal Volume
- 34
- Journal Issue
- 4
- Journal Page Range
- p. 361-366
- ISSN
- 0254-6086
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 48045726
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; CHARGE-COUPLED DEVICES; DESIGN; DIAGRAMS; ELECTRIC CURRENTS; FLUCTUATIONS; GAIN; KHZ RANGE; PERFORMANCE; PHOTONS; READOUT SYSTEMS; SIGNAL-TO-NOISE RATIO; STREAK CAMERAS; VISIBLE RADIATION
- Descriptors DEC
- AMPLIFICATION; BOSONS; CAMERAS; CURRENTS; DIMENSIONLESS NUMBERS; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; FREQUENCY RANGE; INFORMATION; MASSLESS PARTICLES; RADIATIONS; SEMICONDUCTOR DEVICES; VARIATIONS
Optional Information
- Notes
- 5 figs., 11 refs.